Technical Specification
Bolus: 600ml/h
Flow rate range: 1-600ml/h
Volume limit: 1-9999 ml
Accuracy:±5%
KVO flow rate:1ml/h, keep vein open KVO rate
Power supply: AC100-240V, 50/60 Hz, 25VA
Water Proof: IP *I
Battery: Rechargeable lithium polymer battery,7.4 V,1650mAh
Electrical safety: compliance with the requirements of IEC 60601-1.
Max. Power consummation: 25W
Battery recharge: When the pump is connected to the AC power, the battery will automatically recharge About 8-14 hours to recharge fully Can run for more than 5 hours continuously after fully recharged.
Fuse Type:220V 2A*2,12V 2A*2
Display or information: Flow rate, volume limit, accumulated volume, power indicator light, bed No., air, occlusion, empty.
Alarm function: Infusion completion, occlusion, air bubble, low battery, control abnormal, no AC power supply, installation error.
Max. size of outer shell:120*140*190 mm-length* width* height
Max. weight:<2.5 kg
Classification: Type B.
Outer shell material: ABS plastic
operating condition: Environment Environment temperature:+5℃- +40℃,atmosphere pressure 50-106kPa,related humidity 30%-90%
Storage and transport condition: Environment temperature -15℃- +50℃,atmosphere pressure 50-106kPa,relative humidity 30%-90%.
Applicable infusion pipe: All standard infusion pipe use "double dove" to test
EMC: Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.