Feature:
Routine Check Mode and Continuous Monitoring Mode
Data Graph and Trend Table Review
Rich Analysis Report
Perfect Mount Solution
Convenient operation by touch screen
Large SD card memory for more than 2000 patient cases storage
Sync with PC based Sofrware (WinXP and WinVista)
Performance Specifications:
Display:3.5" Color TFT
Resolution:320x240
Display Mode:Standard face, Waveform face, SPO2 view, ECG view, BigNum view
Indicator:Power indicator light, Alarm sound, Pluse tone
Interface:One dual-purpose socket for connecting sensors and communication cables
Power Supply:DC 5V, <500MA
Battery:Builtin Li-Polymer, 2 hours for charging, 3 hours
for continous working.10 hours for standby mode
Trend Graph:Resolution from 1s, 5s, 10s. Maximum time 96 hours.
Trend Table:Resolution from 1s, 5s, 10s. Review up to 1000 items.
Alarm:Ajustable High and Low limits. Three level audible and vistual alarm
User Configuration:All the color of parameters and waves can be set according to user's willing
Technical specifications:
Safety
Meet the requirment of IEC60601 series
Type of Protection:Class II with internal electirc power supply
Degree of Protection:BF
Dimention and Weight
Dimention:92(W)x82(H)x22(D)mm
Weight:136g(with battery)
Operation Environment:
Temperature:0℃~ +40℃
Humidity:15% ~ 95%
Storage Environment:
Temperature:-20℃~ +60℃
Humidity:10% ~ 95%
Patient Range
Neonate ,Pediatric and Adult
SPO2:
Measurement Range:0 ~ 100%
Resolution:1%
Accuracy:±2%(70%~100%, Adult/Pediatric, non-motion)
±3%(70%~100%, Neonate, non-motion)
0% ~ 69% unspecified
Alarm Range:0% ~ 100%
Refreshing Rate:1s
Pulse Rate:
Measurement Range:25 ~ 250 bpm
Resolution:1bpm
Accuracy:±bpm (non-motion)
Alarm Range:25 ~ 250 bmp
Refreshing Rate:1s
ECG:
Lead mode:3-lead or 5-lead
Lead selection: I, II, III, AVR, AVL, AVF, V
HR range:15 ~ 300 bpm
ECG Waveform:2 channals
Accuracy:±2bpm or ±2%, which is greater
S-T segment detection, range: -2.0mv ~ +2.0mv
Arrhythmia analysis: 13 kinds of arrhythmia
Alarm Range:15 ~ 300 bpm
NIBP:
Method:Oscillometry
Mode:Manual/Auto/Contiuous
Measurement Range:25 ~ 260mmHg
Atuo Measure Interval: 5, 10, 15, 30, 45, 60, 90 minutes
Resolution:1mmHg
Overpressure Protection: 300mmHg
Alarm range:25 ~260mmHg